Features for MT-T4A:
1.LCD display can show measuring parameters and results.
2.Pre-store with preselected endpoint titration, blank titration or manual titration function and other automatically generate specific titration mode can be programmable.
3.There are several different titration modes with different sensors.
4.PWM modulation technique stirrer system can be adjusted by software, low noise.
5.With RS232 communication port and can be connected with printer.
6.Anti perchloric acid corrosion material.
Features for MF-T4B:
1. Large LCD screen.
2. Integrated design of valve and burette.
3. Built-in stirrer.
4. Titration transmission device adopts linear motor design.
5. With anti-leakage device.
Technical Parameters:
Model | MF-T4A | MF-T4B |
Measuring Range | pH: 0.~14.00pH mV: -1800.0~1800.0mVTemp.:-5.0~105.0℃ | |
Measuring Resolution | pH: 0.01pH; mV: 0.1mV; Temp:0.1℃ | |
Measuring Accuracy | pH: ±0.01pH±1bit mV: ±0.03%(full scale)Temp.:±0.3℃ | |
Burette Volume Accuracy | 10ml burette:±0.025ml 20ml burette:±0.035ml | |
Burette Resolution | 10ml burette: 1/10000; 20ml burette: 2/10000 | 1/20000 |
Burette Supplementary Speed | 55±10s (when burette is full scale) | |
Repeatability Titration Analysis | 0.2% | |
Repeatability Accuracy | ≤0.2mV | |
Stability | ±0.3 mV±1bit/3h | |
Sensitive of Control Titration | ±2mV | |
Power Supply | AC220V/110V±10%, 50/60Hz | |
Consumption | 45W | |
External Size(W*D*H) mm | 340*400*400 | 390*380*300 |
Package Size(W*D*H)mm | 540*490*480 | |
Gross Weight | 13kg |
Copyrights © Medfuture Biotech Co., Ltd. All Rights Reserved