The ion chromatographic technology is based on the principle of ion exchange and use thin housing ion chromatographic column to quickly separate various ions.
Product Features:
1.The ion chromatographic technology is based on the principle of ion exchange and use thin housing ion chromatographic column to quickly separate various ions. And then, in self-regenerative suppressor, which is at the back of separating column, series connection is used to dissolve strong electrolyte in rinse water in order to deduct its background conductance. And then use conductance detector to continuously detect conductance value of output fluid and we can get chromatographic peak value of various ion. This accomplishes the aim of separating, analyzing qualitatively and quantitatively at one time.
2.MIC-D100 model ion chromatographic meter has the analytic model of suppress conductance detecting anion. We can use different types of chromatographic column to analyze various anions. We can detect not only such conventional anion as F-, CI-, NO2-, PO43-, Br-, NO3-, SO42-, but also CIO2-, CIO3-, BrO3-, etc.
3.At present, HPIC analytic technology has been applied to various fields such as environmental monitoring, sanitation and epidemic prevention, medicines, chemical metallurgy, geology, hydrology, agriculture, forage, electronics industry, electric power, atomic energy, food and drinks, drinking water, scientific research, education, etc.
Technical Parameters:
Model | MIC-D100 | ||
Ion Chromatography Pump | Type | High pressure double plunger tandem advection pump | |
High Pressure Pump Maximum Pressure | 42MPa | ||
Flow Range | 0.001~9.999mL/min | ||
Flow Accuracy | RSD < 0.1 % | ||
Flow Repeatability | RSD≤0.2 % | ||
Conduction Detection | Type | Five electrodes conductance detector | |
Cell Volume | ≤0.8μL | ||
Resolution | ≤0.0020ns/cm | ||
Signal Output Range | 0~35000μS/cm(10 files available) | ||
Output Voltage | -6000~6000mv | ||
Baseline Noise | ≤0.6% FS | ||
Baseline Drift | ≤3% FS | ||
Range of Temperature | RT+5 ℃ ~ 60 ℃ | ||
The Precision of Temperature Control | ±0.01 ℃ | ||
Temperature Compensation | 1.7 %/℃; | ||
Maximum Operating Pressure | 10.0MPa | ||
Flow Path System | Plastic Flow Path | The whole PEEK material | |
Six-way Valve | PEEK material, pressure 5000psi; a signal automatic collecting and flow function. | ||
Analysis Ability | Anion Analysis | F-、Cl-、Br-、I-、NO2-、PO43-、NO3-、SO42-and other anions | |
Cation Analysis | Li+、Na+、NH4+、K+、Mg2+、Ca2+、Sr2+、Ba2+ and other cations | ||
Linear Range | ≥103 | ||
Analysis of Reproducibility | ≤1.0% | ||
Minimum Detectable Concentration | Cl-≤0.0005ug/ml, BrO3-≤0.001ug/ml, Li+≤0.001ug/ml | ||
Power Supply | AC220V±10%, 50Hz | ||
Consumption | 150W | ||
External Size (W*D*H*) mm | 325*400*500 | ||
Package Size (W*D*H*) mm | 625*700*800 | ||
Net Weight | 21Kg | ||
Gross Weight | 28Kg |
Copyrights © Medfuture Biotech Co., Ltd. All Rights Reserved
Technical Support: Sitemap